PRINCIPLE OF OPERATION
Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.
INITIALIZATION TIME
Capnogram within 15 sec., full specifications within 60 sec.
CALIBRATION
Routine calibration is not required. 15 sec. adapter zero performed when changing to different style of adapter.
ACCURACY VERIFICATION
Accuracy verifier provided on sensor cable.
CAPNOSTAT MAINSTREAM CO2 SENSOR
| Size: | 1.30 * 1.67 * 0.85 inches (3.30 * 4.24 * 2.16 cm) |
| Weight: | +/- 1 % of full scale |
| Resolution: | 0.64 oz. (18 grams) - cable excluded |
| Shock Resistance: | Withstands repeated 6 foot drops |
AIRWAY ADAPTERS
Adult/pediatric reusable or single patient use (added dead space <5cc), neonatal reusable (added dead space <.5cc), and sampling adapter for non-intubated patients.
SAMPLING FLOW RATE
180 ml/min (for non-intubated patients).
CO2 (CARBON DIOXIDE)
| Compensations: | N2O & O2 (selectable), BP (automatic) |
| Capnogram: | Selectable sweep speed (12.5 or 25 mm/sec) |
| Range: | 0 - 100 mmHg |
| Accuracy: | ±2 mmHg (for 0 - 40 mmHg) ±5% of reading (for 41 - 70 mmHg) ±8% of reading (for 71 - 100 mmHg) |
| Resolution: | 1 mmHg |
| Response Time: | 60 ms |
RESPIRATORY RATE
| Accuracy: | ±1 br/min |
| Range: | 0-150 br/min |
| Resolution: | 1 br/min |
PRINCIPLE OF OPERATION
Red/Infrared absorption.
SpO2 (OXYGEN SATURATION)
| Range: | 0-100% |
| Accuracy: |
±2% SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2 |
| Resolution: | 1% |
| Averaging: | Menu selectable, 2 or 8 sec. |
| Audio: | Pitch of pulse tone varies with SpO2 value. |
PULSE RATE
| Range: | 30-250 bpm |
| Accuracy: | ±1% of full scale |
| Resolution: | 1 bpm |
| Averaging: | 8 sec. |
SENSORS
Reusable Y-SensorT (can be sterilized and used with all patient populations); reusable adult finger sensor.
PLETHYSMOGRAM
Pulsatile waveform with autogain on/off selection.
ALERTS
| Limits: | Adjustable limits for ETCO2, SpO2, Respiratory and Pulse Rates |
| Audio: | Adjustable volume, 2 min. silence or OFF (LED indicators) |
| Visual: | On-screen & red "Alert Bar" |
DISPLAY
| Type: | Dot matrix, Cold Cathode Display (CCD) |
| Size: | 5" W * 1.5" H (12.7 * 3.08 cm) |
GRAPHIC TREND/HISTOGRAM
| Memory: | 24 hr., battery backed |
| Format: | On-screen or printed, in 30 min., 2, 8, or 12 hr. segments |
COMMUNICATION OUTPUT
| Digital: | RS232 (printer, NovaCARD, computer interface) |
| Analog: | Optional module, 0-1VDC (numerics & waveforms) |
PHYSICAL
| Size: | 3.3" H * 9.0" W * 8.0" D (8.38 * 22.86 * 20.32 cm) |
| Weight: | 8 lb (3.63 kg |
ELECTRICAL
| Power requirements: | 100-120/200-240 VAC, 50-60 Hz, 30 VA |
| Battery: | Sealed lead acid gel cell, 2 hr. life, 12 hr. recharge |
ENVIRONMENTAL
| Operating temperature: | 50-104° F (10-40° C) |
| Operating humidity: | 0-90% relative (non-condensing) |